1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  <20172018  2019  2020  2021  2022  2023  2024  Index 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  <20172018  2019  2020  2021  2022  2023  2024 
<== Date ==> <== Thread ==>

Subject: Re: sumitomo f70 compressor
From: "Lionberger, Carl A." <[email protected]>
To: "[email protected]" <[email protected]>, "[email protected]" <[email protected]>
Date: Tue, 11 Apr 2017 19:18:02 +0000

Hi Linda,


Good information about the firmware versions, thanks.


I'll send you the support off-list.


Carl


Carl Lionberger
Instrument Data Acquisition and Controls
Spallation Neutron Source

From: [email protected] <[email protected]> on behalf of [email protected] <[email protected]>
Sent: Tuesday, April 11, 2017 2:38 PM
To: [email protected]
Subject: RE: sumitomo f70 compressor
 

Pierrick,

 

Watch out for firmware version and make sure your manual version matches your firmware version and the equipment because the same controller is used for slightly different applications pre-configured.   We have two F-70H we acquired a while ago with supply of instrumentation for an End Station.   

The manual did not describe monitoring via serial link so we tried to control and monitor them via VME digital IO and analogue cards.

 

It turns out that the original firmware did not support serial monitoring and control, but later versions of the firmware do.  We actually have two Sumitomos with the same model number but with different versions of the firmware V1-5(not serial capable) and V1-7 (serial capable – v1-6 and above) and two different versions of the manual were supplied at different times.   The firmware version can be displayed on the little screen.  (The outcome was that we achieved some monitoring, but not control because the system can be used in different “modes” based on the “mode configuration” switch which affects what the pins mean and it was wired for one mode and needed to be operated in another.   However staff on the beamline owning the compressors were satisfied with the monitoring).

 

Carl,

 

Another of our beamlines now has Sumitomo compressors with newer firmware and they use CSS, so we would be interested in the serial support for that, is there a link for it?

 

Thanks,

 

Linda Pratt

Software Systems Engineer

Controls Department

[email protected]

+44 (0)1235 778058

www.diamond.ac.uk

Diamond Light Source Ltd

Diamond House

Harwell Science and Innovation Campus

DIDCOT

Oxfordshire

OX11 0DE

UK

 

From: [email protected] [mailto:[email protected]] On Behalf Of Lionberger, Carl A.
Sent: 14 February 2017 18:08
To: Pierrick Hanlet; [email protected]
Subject: Re: sumitomo f70 compressor

 

We have support that uses the serial port, with stream device.  The GUI is for BOY.  

 

Carl

 

Carl Lionberger
Instrument Data Acquisition and Controls
Spallation Neutron Source


From: [email protected] <[email protected]> on behalf of Pierrick Hanlet <[email protected]>
Sent: Tuesday, February 14, 2017 12:22 PM
To: [email protected]
Subject: sumitomo f70 compressor

 

Hi,
Does anyone have support for a Sumitomo F70 cryo-compressor?
Thank you,
Pierrick

--
"Whether you think you can or think you can't, either way, you are correct" -- Henry Ford
_______________________________________________________________
Pierrick Hanlet
IIT/Fermilab
+44-(0)1235-44-5550 (RAL)
+1-630-840-5555 (FNAL)
+1-312-567-5745 (IIT)
+1-630-888-0185/+44-(0)79-48-860-197 (US/UK mobile)

 

-- 

This e-mail and any attachments may contain confidential, copyright and or privileged material, and are for the use of the intended addressee only. If you are not the intended addressee or an authorised recipient of the addressee please notify us of receipt by returning the e-mail and do not use, copy, retain, distribute or disclose the information in or attached to the e-mail.
Any opinions expressed within this e-mail are those of the individual and not necessarily of Diamond Light Source Ltd.
Diamond Light Source Ltd. cannot guarantee that this e-mail or any attachments are free from viruses and we cannot accept liability for any damage which you may sustain as a result of software viruses which may be transmitted in or with the message.
Diamond Light Source Limited (company no. 4375679). Registered in England and Wales with its registered office at Diamond House, Harwell Science and Innovation Campus, Didcot, Oxfordshire, OX11 0DE, United Kingdom
 


References:
sumitomo f70 compressor Pierrick Hanlet
Re: sumitomo f70 compressor Lionberger, Carl A.
RE: sumitomo f70 compressor Linda.Pratt

Navigate by Date:
Prev: Re: [AsynPortDriver] "Multiple interrupt callbacks between processing" warning with multiple arrays Paweł Plewiński
Next: RE: [AsynPortDriver] "Multiple interrupt callbacks between processing" warning with multiple arrays Mark Rivers
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  <20172018  2019  2020  2021  2022  2023  2024 
Navigate by Thread:
Prev: RE: sumitomo f70 compressor Linda.Pratt
Next: Tech-talk archives & RSS Andrew Johnson
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  <20172018  2019  2020  2021  2022  2023  2024 
ANJ, 21 Dec 2017 Valid HTML 4.01! · Home · News · About · Base · Modules · Extensions · Distributions · Download ·
· Search · EPICS V4 · IRMIS · Talk · Bugs · Documents · Links · Licensing ·