EPICS Controls Argonne National Laboratory

Experimental Physics and
Industrial Control System

1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  <20142015  2016  2017  2018  2019  2020  2021  2022  2023  2024  Index 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  <20142015  2016  2017  2018  2019  2020  2021  2022  2023  2024 
<== Date ==> <== Thread ==>

Subject: RE: Keithley 2400 communication
From: "Mooney, Tim M." <[email protected]>
To: priya tiwari <[email protected]>, "[email protected]" <[email protected]>
Date: Thu, 11 Sep 2014 13:48:48 +0000
Hi Priya,

A quick way to do this would be to attach an asyn record to the serial port, and use a sseq record to control and send strings to it.

record(sseq, "doReset") {
field(STR1, "Read")
field(LNK1, "asynrecord.TMOD")

field(STR2, "*RST")
field(LNK2, "asynrecord.AOUT CA")
field(WAIT2, "Wait")

field(STR3, ":SOUR:FUNC VOLT")
field(LNK3, "asynrecord.AOUT CA")
field(WAIT3, "Wait")
...
field(STR7, ":OUTP ON")
field(LNK7, "asynrecord.AOUT CA")
field(WAIT7, "Wait")

field(STR8, "Write/Read")
field(LNK8, "asynrecord.TMOD")

field(STR9, ":READ?")
field(LNK9, "asynrecord.AOUT CA")
field(WAIT9, "Wait")
}

Tim Mooney ([email protected]) (630)252-5417
Software Services Group (www.aps.anl.gov)
Advanced Photon Source, Argonne National Lab


From: [email protected] [[email protected]] on behalf of priya tiwari [[email protected]]
Sent: Thursday, September 11, 2014 2:19 AM
To: [email protected]
Subject: Keithley 2400 communication

Hello,
        I am trying communication with KEITHLEY 2400 on serial port(RS-232).I sent command "*IDN?" it has performed very well and i got the output.
i want to reset the device and the commands are
*RST
:SOUR:FUNC VOLT
:SOUR:VOLT:MODE FIX
:SOUR:VOLT:RANG 20
:SOUR:VOLT:LEV 10
:OUTP ON
:READ?
now,How to write program in VDCT for performing all commands sequentially.Actually i have no idea about this.please help me....

Q.2: How to send commands to device  at run time in VDCT?
what is the format of
caput? on that time


Thanks in advance
    

Replies:
Re: Keithley 2400 communication Rod Nussbaumer
References:
Keithley 2400 communication priya tiwari

Navigate by Date:
Prev: Re: msi missing when isntalling synApps Mark Rivers
Next: Re: Keithley 2400 communication Rod Nussbaumer
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  <20142015  2016  2017  2018  2019  2020  2021  2022  2023  2024 
Navigate by Thread:
Prev: RE: Keithley 2400 communication nick.rees
Next: Re: Keithley 2400 communication Rod Nussbaumer
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  <20142015  2016  2017  2018  2019  2020  2021  2022  2023  2024 