It sounds like it could be a problem with EPICS_CA_MAX_ARRAY_BYTES. That must be set to a value at least slightly larger than the largest image size you want to collect. The number of elements in your waveform record needs to be at least equal to the number of pixels in the camera.
EPICS_CA_MAX_ARRAY_BYTES must be set on both the IOC and CA client computers.
What client are you using?
Mark
________________________________________
From: [email protected] [[email protected]] on behalf of [email protected] [[email protected]]
Sent: Friday, April 26, 2013 9:18 AM
To: [email protected]
Subject: Unable to get a picture from Baumer TXG13 GigE camera
Hi all,
We're trying to set up a Baumer TXG13 monochrome CCD camera using aravisGigE. EPICS connects successfully with the camera and we can read and set the parameters. When setting AcquisitionStart=1 the ArrayCounter starts increasing (but not the FrameCounter). With tcpdump we see also the image data is transmitted and we also see higher or lower values with or without lens cover and also increasing numbers with the test image enabled. Unfortunately we never where able to retrieve the image data from NDArray, ArrayData always delivers an array filled with 0s. We are not sure now if the problem is related with the camera configuration or with NDArray. Is it normal that the FrameCounter remains 0? In case the camera is properly configured, we will be missing some configuration of NDArray. We tried EnableCallbacks=1 and also tried setting the array dimensions according to the resolution 1392x1040, Mono8. We might be missing something very basic, any suggestions and example configur!
ations are warmly welcome.
Regards,
Urs
_____________________________________
University of Bern
Institute of Physics
Space Research and Planetary Sciences
Urs Geissbühler
Sidlerstrasse 5
CH-3012 Bern
Direct +41 (0)31 631 4425
Main +41 (0)31 631 4411
Fax +41 (0)31 631 4405
[email protected]
http://space.unibe.ch/
- References:
- Unable to get a picture from Baumer TXG13 GigE camera urs.geissbuehler
- Navigate by Date:
- Prev:
RE: Unable to get a picture from Baumer TXG13 GigE camera tom.cobb
- Next:
Re: the cheksum of Streamdevice Dirk Zimoch
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
2009
2010
2011
2012
<2013>
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
- Navigate by Thread:
- Prev:
RE: Unable to get a picture from Baumer TXG13 GigE camera tom.cobb
- Next:
RE: Unable to get a picture from Baumer TXG13 GigE camera urs.geissbuehler
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
2009
2010
2011
2012
<2013>
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
|