Hello Luo,
Did you confirm the IP or VXI-11 connection from your Linux PC to
the SL1000?
Your log message contains "#!../../bin/solaris-sparc/scopetest".
Is it OK?
Which version of sl1000 device support do you use? Do you use the
sl100 firmware version later than Ver2.03B005?
Kazuro.
P.S. Is it correct to use <[email protected]> as the
mail-list address?
>>> On Wed, 15 Jul 2009 17:49:15 JST, "Luo, Xizhi" <[email protected]> wrote;
> Hello, everyone
>
>
> I am a beginner to use EPICS system.
> I want to connect to an instrument SL1000 via VXI11 mode.
> But unluckily some error happened.
>
> My PC environment:
> OS: CentOS5.3
> EPICS_BASE: base-3.14.10
> asynDriver: asyn-4.11
> instrument: sl1000-1.1
>
> So someone could tell why connect is failed?
> And what is the problem with my epics project?
> Thanks very much!
>
> Error message is followed:
> [root@luoxizhi-01 iocscope]# ../../bin/linux-x86/scope st.cmd
> #!../../bin/solaris-sparc/scopetest
> < envPaths
> epicsEnvSet("ARCH","linux-x86")
> epicsEnvSet("IOC","iocscope")
> epicsEnvSet("TOP","/opt/epics/modules/sl1000-1.1")
> epicsEnvSet("SUPPORT","/opt/epics")
> epicsEnvSet("ASYN","/opt/epics/modules/asyn")
> epicsEnvSet("EPICS_BASE","/opt/epics/base")
> < envParams
> epicsEnvSet EPICS_CA_MAX_ARRAY_BYTES 5000000
> #
> # Site-specific configuration
> #
> epicsEnvSet(ETHER1, "10.0.158.39") # Choose scope ethernet address
> epicsEnvSet(SCOPE1, "scope1:") # Choose scope name
> #epicsEnvSet(ETHER2, "192.168.21.55") # Choose scope ethernet address
> #epicsEnvSet(SCOPE2, "scope2:") # Choose scope name
> #
> # Ensure that we can transfer a complete waveform
> #
> epicsEnvSet(EPICS_CA_MAX_ARRAY_BYTES, 5000000)
> ## Register all support components
> cd /opt/epics/modules/sl1000-1.1
> dbLoadDatabase("dbd/scope.dbd")
> scope_registerRecordDeviceDriver(pdbbase)
> ## This is the main db.
> dbLoadRecords("db/SL1000.db","scope=scope1:,L=0,A=0,ASG=TEST1,Name=YOKOG
> AWA SL1000-1,WF_NELM=10000,DISPWF_NELM=1000,SRQ_EVNT=101")
> #dbLoadRecords("db/SL1000.db","scope=$(SCOPE2),L=0,A=0,ASG=TEST2,Name=YO
> KOGAWA SL1000-2,WF_NELM=10000,DISPWF_NELM=1000,SRQ_EVNT=102")
> ## Set up IOC/hardware links -- Direct network connection
> vxi11Configure("L0", "10.0.158.39", 0, 0.0, "inst0", 0)
> #vxi11Configure("L1", "$(ETHER2)", 0, 0.0, "inst0", 0)
> ## Start IOC
> cd /opt/epics/modules/sl1000-1.1/iocBoot/iocscope
> iocInit()
> Starting iocInit
> ########################################################################
> ####
> ## EPICS R3.14.10 $R3-14-10$ $2008/10/27 19:39:04$
> ## EPICS Base built Jul 15 2009
> ########################################################################
> ####
> 2009/07/16 01:37:17.577 L0 vxi11 clientCall errno Success clnt_stat 5
> 2009/07/16 01:37:17.577 L0 vxiCreateIrqChannel (create_intr_chan): RPC:
> Timed out
>
> 2009/07/16 01:37:17.577 L0 Warning -- SRQ not operational.
> Segmentation fault
> [root@luoxizhi-01 iocscope]#
-----
Kazuro FURUKAWA <[email protected]>
Linac&KEKB, High Energy Accelerator Research Organization (KEK), Japan
Telephone: +81-29-864-5200 x4316, Facsimile: +81-29-864-0321
- References:
- VXI11 connect failed Luo, Xizhi
- Navigate by Date:
- Prev:
RE: When an external device does not provide termination on TCP ASCIIstream replies Allison, Stephanie
- Next:
c channel access with functions Patrick Thomas
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
<2009>
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
- Navigate by Thread:
- Prev:
VXI11 connect failed Luo, Xizhi
- Next:
uptodate version of modtcp? Burkhard Kolb
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
<2009>
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
|