If you mean, communicating over the serial line, yes we have at LCLS,
using streamdevice. If you like, I can send you:
streamdevice protocol file
database to go with it
edm screen files
We use the identical stuff for the Varian 701, 1001, and V81.
Note that we have 3x V81, 2x 301, 2x 701, 2x 1001 but of the 9 units,
only the 3x V81 and one of the 301 communicate reliably. The other 6
simply stop communicating within 20-60 minutes; only a power cycle
recovers it. Since we use a PLC connected to the voltage signals for
critical control and monitoring, we can survive without damage. I
have greatly increased all the time-delays to account for
manufacturing differences, to no avail. I suspect that the firmware
is bad.
At 10:23 AM -0800 2009/03/06, Dave Reid wrote:
Has anyone interfaced the Varian 301-AG turbo pump controllers to EPICS?
Thanks...
dave
- References:
- Varian Turbo Pump Controllers Dave Reid
- Navigate by Date:
- Prev:
Varian Turbo Pump Controllers Dave Reid
- Next:
EPICS for USB devices Mehta, Pratik
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
<2009>
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
- Navigate by Thread:
- Prev:
Varian Turbo Pump Controllers Dave Reid
- Next:
EPICS for USB devices Mehta, Pratik
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
<2009>
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
|