1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  <20072008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024  Index 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  <20072008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024 
<== Date ==> <== Thread ==>

Subject: Re: current measurement - hardware
From: "David Dudley" <[email protected]>
To: <[email protected]>
Date: Mon, 26 Nov 2007 11:40:53 -0600
Might look into something like a Measurement Computing PCI2517 card,
which has 8 differential analog inputs.

Looking at the spec sheet, it says it has a 40pA bias current, and
20Meg input impedance, while measuring a 100mv signal.  With proper load
resistors (if you even need them at these low currents), you could
probably measure currents in the Ua range with no problem.

David Dudley

>>> Heinrich du Toit <[email protected]> 11/26/2007 8:25 AM >>>
Hi guys

I'm looking for some information on how you solve this problem...

We need to measure currents.. from harps and slits and faraday cups.

All of this gives of currents that is important.
Some of the currents we measure with keithleys but it is way to 
expensive to buy keithleys for harps as you will need 100's of them :)

We need to measure currents between nA and uA range for harps and
slits.
For faraday cups this sometimes goes to 100+ uA but our main problem is

harps/slits.

I hope that we are not the first to have this problem, is there some 
cost effective method to solve this problem? That can give descent
results.

The problem can basically be given as:
How do I measure 50+ current inputs (harp = 96, 4 per slit-set) across
a 
wide range (2nA fullscale = 50pA sensitive up to 100uA full scale)
still 
have resonable accuracy especially between the lines and have a descent

update rate and not 2 high noise levels.
(I guess noise can to some extend be solved with a software filter if 
updates are high enough)

Thanks for your support :)
-Heinrich



BEGIN:VCARD
VERSION:2.1
X-GWTYPE:USER
FN:David Dudley
TEL;WORK:880-3740
ORG:;MIS
TEL;PREF;FAX:880-3741
EMAIL;WORK;PREF;NGW:[email protected]
N:Dudley;David
END:VCARD


Replies:
Re: current measurement - hardware Mark Davis
References:
current measurement - hardware Heinrich du Toit

Navigate by Date:
Prev: Re: current measurement - hardware Tim Mooney
Next: EPICS PHP Extension David Dudley
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  <20072008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024 
Navigate by Thread:
Prev: Re: current measurement - hardware Tim Mooney
Next: Re: current measurement - hardware Mark Davis
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  <20072008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024 
ANJ, 10 Nov 2011 Valid HTML 4.01! · Home · News · About · Base · Modules · Extensions · Distributions · Download ·
· Search · EPICS V4 · IRMIS · Talk · Bugs · Documents · Links · Licensing ·