On Thu, Dec 15, 2005 at 09:08:19AM -0600, Chithra Karunakaran wrote:
> Hi:
>
> Has anyone worked with Oxford Danfysik Beam Position monitoring system
> (4 channel IC plus Electronics Unit, YMCS0012). The device does not
> respond to any of the commands send through the serial port.
>
> Thanks,
> Chithra.
Hi. I have.
I am able to communicate with it with the following RS-232 settings:
Baud rate: 19200
Data format: 8 bits
Stop bits: 1
Parity: none
Flow control: none
Also, if you have only one device (i.e. you are not chaining the
devices), your RS-232 cable must be wired correctly. The wiring should
be as follows:
VME <-> QBPM
--- ----
2 3
3 2
5 5
4-6-8
On the end that connects to the VME IOC, pins 4, 6, and 8 should be tied
together.
If you have more than one device, the wiring needs to be different (see
section 6.6 on page 13 of the manual).
A command you can send is ":READ1:CURRALL?" terminated with "\n" (ASCII
10 / linefeed) to read all four current channels at once. Replace the 1
after "READ" with the unit address assigned to your device. Or if you
don't know it, try 0-15 if 1 doesn't work.
Hope this helps.
-lewis
--
J. Lewis Muir
Software Engineer
CARS/IMCA-CAT
- References:
- Oxford Danfysik-YMCS0012 Chithra Karunakaran
- Navigate by Date:
- Prev:
Oxford Danfysik-YMCS0012 Chithra Karunakaran
- Next:
RE: about CA monitor Jeff Hill
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
<2005>
2006
2007
2008
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
- Navigate by Thread:
- Prev:
Oxford Danfysik-YMCS0012 Chithra Karunakaran
- Next:
RE: Oxford Danfysik-YMCS0012 Walters, M (Malcolm)
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
<2005>
2006
2007
2008
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
|