Beamline 5-ID-B,C,D: Sector 5 - Insertion Device Beamline
DND-CAT
Materials Science, Polymer Science
Description
The 5-ID beamline contains three end stations for use by general users.
The 5-ID-B station contains a custom 11-detector powder camera formerly stationed at COM-CAT, 32-ID. This system has 11 Si(111) analyzer crystals and operates at an energy of ~ 17.5 keV for high-resolution powder diffraction measurements. The system has a six-position sample changer for capillaries.
The 5-ID-C station has two instruments, a kappa diffractometer for in-air or small sample environment cells, and a UHV chamber with MBE and other capabilities. This station has post-monochromator optics for beam conditioning primarily for x-ray standing wave measurements. The UHV system and the post-monochromator optics are only available to general users in a collaborative mode.
The 5-ID-D station contains a pinhole small-angle x-ray scattering (SAXS) camera with a maximum sample-to-detector distance of approximately 10 m. Multiple samples can be loaded onto sample wheels and placed inside the vacuum system to produce ultra-low-background SAXS patterns. Simultaneous small- and wide-angle diffraction (SAXS/WAXS) with custom Roper CCD detectors is available on 5-ID-D, including an optional in-vacuum flow cell for solution SAXS/WAXS.
|
 |
 
Supported Techniques
- Powder diffraction
- X-ray standing waves
- X-ray optics development/techniques
- Small-angle x-ray scattering
- Surface diffraction
- X-ray reflectivity
- Wide-angle x-ray scattering
Beamline Controls and Data Acquisition
Linux running custom control code for motors and actuators on all systems. SPEC used to control diffractometers. Vendor-supplied software used to control CCD detectors.
Detectors
- Scintillation detectors
- Mar 165 mm CCD detector
- Microphotonics GEMSTAR 2-D detector
- Roper SAXS/WAXS system
- 11-detector powder camera
- Vortex Silicon Drift Detector
Additional Equipment
- Small kappa 5-circle diffractometer
- Capillary lens
- In-vacuum SAXS setup
- Simultaneous SAXS/WAXS
- Instron servo-hydraulic tensile stage (collaboration only)
- Solution SAXS in-vacuum flow cell
|
|
Local Contacts
Beamline Specs
Source |
3.3 Undulator (Undulator A) |
Monochromator Type |
Si(111) |
Energy Range |
5-20 keV |
Resolution (ΔE/E) |
|
Flux (photons/sec) |
@ keV |
Beam Size (HxV) |
  |
Focused |
300µm x 1000µm
|
Unfocused |
2mm x 1mm
|
For additional information see:
http://www.dnd.aps.anl.gov/
|