Beamline 23-ID-D: Sector 23 - Insertion Device Beamline D

GM/CA-CAT
Life Sciences

Description

Sector 23 utilizes canted undulators to provide two independent sources for the two ID-beamlines 23-ID-B (U33) and 23-ID-D (U30). The beamlines support macromolecular crystallography, MAD/SAD phasing, virus crystallography, and microcrystals. The beamlines are equipped with on-axis-microscopes allowing parallax free alignment. Our quad-mini-beam collimators (Fischetti, et.al. JSR 16, 217-225) provide either a 5, 10 or 20 micron beam or a scatter guard for the 70x25 micron^2 beam with intensities of 2.4e11, 6.2e11, 2.4e12 and 1.3e13 photons/sec, respectively. The sample goniometer has a ~1 micron (peak-to-peak) sphere-of-confusion. The mini-beam capability has been coupled with powerful user friendly software (JBluIce-EPICS) providing the ability to raster scan the diffractive or fluorescence properties of a sample to locate micro-crystals or map the properties of larger crystals. Users can also mitigate the effects of radiation damage by defining a 3-dimensional vector along which partial data sets can automatically be collected. An ALS-style sample automounter (BAM-1) provides high throughput sample screening. The automounter accommodates ALS-Hampton pins in ALS, Unipuck, or Rigagu pucks.

 

Supported Techniques

  • Macromolecular crystallography
  • Microbeam
  • Large unit cell crystallography
  • Subatomic (<0.85 Å) resolution
  • Multi-wavelength anomalous dispersion
  • Single-wavelength anomalous dispersion

Beamline Controls and Data Acquisition

The GM/CA JBluIce-EPICS user friendly interface retains the look of the original SSRL BluIce, but has been completely rewritten in JAVA at GM/CA to run on EPICS, VME EPICS/MEDM low level interface, PMAC Turbo motion control, and Linux workstations for data collection and processing (Stepanov, et.al. Acta. Cryst. D. 67, 176-188). The JBluIce-EPICS software is an open source available for downloading from GM/CA-CAT files depository.

Detectors

  • MAR 300 CCD
  • KETEK SDD fluorescence spectroscopy

Additional Equipment

  • Mini-beams: 5, 10 or 20 microns
  • Sample automounter
  • On-axis-visualizer

Local Contacts

Name ROBERT F. FISCHETTI (Associate Director)
Phone 630.252.0660
Name CRAIG OGATA (Crystallography)
Phone 630.252.0675
Name SHEILA Y. TRZNADEL (Scheduling)
Phone 630.252.0662

Beamline Specs

Source

3.0 Undulator

Monochromator Type

Si(111)

Energy Range

5-20 keV

Resolution (ΔE/E)

2 x 10 -4

Flux (photons/sec)

1.3 x 10 13 @12 keV

Beam Size (HxV)

 

Focused

70µm x 25µm

For additional information see:
http://www.gmca.anl.gov/

Current Status:

Operational/Accepting General Users

Access Mode: