DISSOLVED OXYGEN SENSOR MAINTENANCE RECORDS
Service Date Sensor S/N Instrument S/N Location of instrument Work performed Comments
8/13/2004 54228 24999 PAR mezzanine (APS-PW) Reconditioned sensor (Sec. 3 sensor manual) (no NH3/HNO3 used)  
8/13/2004 29140 22978 PAR mezzanine (APS-PW) Reconditioned sensor (Sec. 3 sensor manual) (no NH3/HNO3 used) erratic reading
8/13/2004 41319 25003 Bldg 450 (PW-A) Reconditioned sensor (Sec. 3 sensor manual) (no NH3/HNO3 used)  
8/13/2004 37957 23335 Bldg 450 (PW-B) Reconditioned sensor (Sec. 3 sensor manual) (no NH3/HNO3 used)  
8/13/2004 41320 25008 User water station (mez. col. 71.5) Reconditioned sensor (Sec. 3 sensor manual) (no NH3/HNO3 used)  
8/13/2004 41324 22902 Bldg 450 (Degassifier outlet) Reconditioned sensor (Sec. 3 sensor manual) (no NH3/HNO3 used)  
8/19/2004 41325 25005 PAR mezzanine (PW-A) Reconditioned sensor and replaced instrument Instrument 22978 found to have unstable; sensor 54228 still OK
8/19/2004 29140 24999 PAR mezzanine (PW-B) Sensor connected to different instrument only  
           
           
           
           
           
           
Gray field indicates a more recent entry for sensor/instrument exists